Biometry and IOL power: how formulas model the eye
How the eye is measured, how thin- and thick-lens models turn those measurements into an IOL power, how current formulas add regression and machine learning, and what changes for toric lenses and special eyes.
Measuring the eye
Every formula is only as good as the axial length and corneal power it is given, and each measurement belongs with a matching lens constant.
Axial length
How each biometer measures axial length
- Where each method stops. Optical biometry measures to the RPE, roughly one retinal thickness behind the ultrasound endpoint.
- Speed of sound. A single average speed (about 1550 m/s) misreads eyes whose lens and vitreous proportions differ from average.
- Contact compression. Applanation shortens the reading by a variable amount; one series read 23.64 mm by applanation against 24.20 mm optically.
- Dense cataract. Raises the lens sound speed and can block the optical signal, which is often why an eye ends up on ultrasound.
| Method | Measures to | Constant |
|---|---|---|
| Contact A-scan | ILM, probe can indent cornea | Contact ultrasound constant |
| Immersion A-scan | ILM, fluid separates probe from cornea | Immersion ultrasound constant |
| Original IOLMaster (PCI) | RPE, reading calibrated against immersion | Optical (ULIB / IOLCon) constants |
| Optical, segmented | RPE, index per segment | Segmented-AL constants |
Optical biometry
Why the original IOLMaster was calibrated against immersion ultrasound
Fercher and Roth reported optical measurements of human eye length in 1986. In 1993, Hitzenberger and colleagues compared the methods in cataractous eyes. Their optical readings averaged 0.18 mm longer than immersion ultrasound and 0.47 mm longer than contact ultrasound. Light reflected from the retinal pigment epithelium, deeper than the inner retinal surface that returned the ultrasound echo. With contact ultrasound, pressure from the probe could shorten the reading further.
To use partial coherence interferometry (PCI) with IOL formulas developed around ultrasound, Wolfgang Haigis and colleagues in Würzburg measured the same eyes with a Zeiss prototype and a Grieshaber immersion A-scan. They used the paired readings to convert optical path length into the axial length an immersion scan would give on average. The IOLMaster already applies that conversion to the number on screen. B. A. M. Lege and Haigis reported the prototype validation in 1999. The group’s 2000 paper compared five conversions, including the one Zeiss had built into the instrument.
Displayed AL = (OPL / 1.3549 − 1.3033) / 0.9571
OPL is the optical path length measured by PCI, in millimetres. The prototype divided OPL by 1.3549 to put it on the scale of eye length. The other two coefficients came from the comparison with immersion ultrasound. For example, an OPL of 32.89 mm gives a displayed AL of 24.00 mm. The displayed value is what goes into the IOL formula. Haigis’s derivation, equations 1–3 (PDF).
Warren Hill described the Würzburg calibration in 2003. His own 2008 study with Raymund Angeles and Todd Otani tested whether version 5 software could recover a usable axial-length signal in eyes that earlier software struggled to measure.
Immersion to optical
Using an optical lens constant with an immersion measurement
Calibrating the biometer sets the axial length (AL) reported in millimetres. Optimising a lens constant uses postoperative refractions to tune a formula for a particular IOL. When moving from ultrasound to optical biometry, Haigis suggested a starting SRK/T constant based on the mean difference across paired eyes, with AL in millimetres: A(optical) = A(ultrasound) + 3 × (mean optical AL − mean ultrasound AL).
Cooke and colleagues studied the reverse problem in 2022. If only an immersion reading is available, how closely does it match the scale used with an optical constant? Across 1,970 paired eyes, optical AL averaged 0.087 mm longer. They evaluated two alternatives for their data. One kept the measured immersion AL and subtracted 0.23 from the optical A-constant. The other kept the optical constant and converted the input to 1.0228 × immersion AL − 0.4556 mm. The graphs show why these approaches give similar results near 24 mm and diverge in shorter and longer eyes.
Computed with SRK/T (K 43.5 D, A 119.0). The constant shift removes about 0.5 D in short eyes, where the scales already agree, and almost nothing at 30 mm, where about 0.56 D is needed.
Immersion and SRK/T
SRK/T on immersion: errors can stack or hide
Short eyes
The constant shift moves short eyes 0.5 D where no correction is needed. SRK/T’s own short-eye bias varies in direction: myopic in a 2024 optical study, hyperopic in an older ultrasound series. SRK/T is less accurate here either way (MAE 0.75 D against 0.46 D for Hoffer Q).
Long eyes
The short immersion reading pushes about −0.3 D myopic at 30 mm and hides part of SRK/T’s hyperopic long-eye error (about +0.9 D).
Adjustments
Patching older formulas: what each adjustment fixes
Older formulas were built on simplifying assumptions: one average refractive index for the whole eye, lens position predicted from AL and K alone, and a cornea with a normal back-to-front ratio. Each assumption fails in a particular group of eyes. Rather than replace the formula, a published adjustment corrects one input or one step, and each adjustment fixes only the problem it was designed for.
| Problem | Why it happens | Adjustment | Used with |
|---|---|---|---|
| Axial length on the wrong scale in long and short eyes | A single average refractive index; long eyes have proportionally more vitreous, short eyes more lens | CMAL (Cooke 2019): 1.23853 + 0.95855·AL − 0.05467·LT, approximating sum-of-segments AL; or a sum-of-segments biometer | The AL input, for any formula |
| Hyperopic surprise in long eyes | AL- and K-based lens-position prediction does not extrapolate to long eyes | Wang-Koch AL adjustment (2011), one equation per formula, e.g. SRK/T: AL′ = 0.8544·AL + 3.7222 for AL > 25 mm; later modified versions start at 26.5–27 mm | SRK/T, Holladay 1, Haigis, Hoffer Q |
| Erratic SRK/T results at steep or flat K | SRK/T’s corneal-height step behaves non-physiologically at extremes | T2 formula (Sheard, Smith and Cooke 2010): a regression replaces the corneal-height calculation | SRK/T |
| Overestimated corneal power after myopic LASIK/PRK | The front surface is flattened but the back is not, so the standard ratio no longer holds; formulas that use K for lens position are misled twice | Haigis-L: a correction curve within Haigis, which predicts lens position without K; double-K (Aramberri 2003): pre-surgery K for lens position, post-surgery K for power | Haigis; double-K with SRK/T, Holladay 1, Hoffer Q |
Adjustments are formula-specific and should not be stacked unless a method says so. Current formulas build several of these fixes in: EVO, PEARL-DGS and Cooke K6 use a CMAL-type AL adjustment, and post-refractive versions such as Barrett True-K handle the cornea directly. In one comparison, SRK/T with CMAL performed comparably to newer formulas and better than with the Wang-Koch adjustment.
Keratometry
The 1.3375 index is a convention
| Example cornea (r 7.7 / 6.4 mm) | D |
|---|---|
| Front surface (n 1.376) | +48.8 |
| Back surface (1.336 − 1.376) | −6.3 |
| Net power, with thickness | ≈ 42.7 |
| Keratometric K = 337.5 / r | 43.8 |
Keratometers measure the front radius and assume a normal back surface. The offset between K and true net power is absorbed into each formula’s constants. It breaks when the back-to-front ratio changes, after laser refractive surgery and in keratoconus.
Posterior cornea
K, PK, TK and TCRP
- K: front surface only, with the 1.3375 index.
- PK: posterior keratometry measured by Scheimpflug or OCT, about −6 D in normal eyes.
- TK: total keratometry (IOLMaster 700): front, back and thickness on the keratometric scale.
- TCRP: Pentacam ray-traced total power over a chosen zone.
| Eyes | Evidence |
|---|---|
| Normal, monofocal | K ≈ TK (447 eyes, four formulas) |
| Toric | Mixed; TCRP less accurate in one 23-eye study |
| Multifocal | K slightly better than TK (one study) |
| After myopic LASIK/PRK | TK and PK help |
| Keratoconus | TK and PK help |
Thin-lens formulas
Classic formulas treat the cornea and IOL as single planes. Everything then depends on predicting where the lens will sit.
Cornea and IOL as single planes
P = n / (AL − ELP) − n / (n/K − ELP)
n = 1.336. ELP, the effective lens position, is not the physical IOL position: it is whatever position makes this equation give the right answer.
What sets the power vs what causes the error
Left: thin-lens vergence on a textbook eye (AL 23.5 mm, K 43.5 D, ACD 3.1 mm, LT 4.5 mm), lens position = ACD + 0.4 × LT, one input varied at a time; as shares of variance AL ≈ 83%, K ≈ 15%, ACD ≈ 2%, LT < 1%. AL and K set the power; ACD and LT matter because they predict lens position, the largest source of error.
Generations of lens-position prediction
- SRK / SRK IIRegression only: P = A − 2.5·AL − 0.9·K
- SRK/T, Holladay 1, Hoffer QVergence; ELP from AL and K
- HaigisELP = a0 + a1·ACD + a2·AL, three constants optimised on outcomes
- Holladay 2Seven inputs including LT, WTW and age
- Current formulasThick-lens optics, regression and machine learning
What a lens constant can and cannot do
A lens constant moves the predicted lens position, so it shifts the predicted refraction of every eye by a similar amount. Optimisation back-calculates the constant each operated eye would have needed and averages them, so the mean error becomes zero.
It moves the centre of the error distribution, not its spread, and it cannot correct errors that change with axial length.
Optimised constants: ULIB, IOLcon and each calculator
A manufacturer publishes a nominal A-constant, often intended for contact ultrasound biometry. Outcome data refine it.
ULIB, the User Group for Laser Interference Biometry run by Wolfgang Haigis, collected outcomes with optical biometry and published optimised constants. It has not been updated since 2016, and most of its data moved to IOLcon.
IOLcon (iolcon.org), run by Achim Langenbucher’s group, is ULIB’s successor. It is free, collects outcomes from surgeons, and publishes optimised constants for published formulas, with the number of eyes behind each. It cannot optimise proprietary formulas such as Barrett.
| Lens | Manufacturer nominal | Optical SRK/T (source) | IOLcon optimised SRK/T (eyes) | Barrett (equiv. A) | Kane | Cooke K6 | PEARL-DGS | EVO |
|---|---|---|---|---|---|---|---|---|
| Alcon SN60WF | 118.7 US | 119.0 ULIB | 119.02 (851) | 118.99 | 118.98 | 118.96 | 119.10 | 119.0 |
| Alcon SA60AT | 118.4 US | 118.8 ULIB | 118.85 (1,481) | 118.53 | 118.70 | 118.80 | 118.85 | 118.8 |
| Alcon MA60AC | 118.8 US | 119.2 ULIB | – | 119.20 | 119.24 | 118.90 | – | 119.2 |
| Alcon Clareon CNA0T0 | 118.8 US | 119.1 mfr | 119.35 (1,057) | – | – | – | – | 119.2 |
| J&J Tecnis ZCB00 | 119.3 | 119.3 mfr | 119.41 (1,670) | 119.39 | 119.36 | 119.43 | 119.35 | 119.3 |
| J&J Tecnis Eyhance | 119.3 | 119.3 mfr | 119.26 (134) | 119.39 | – | 119.43 | – | 119.3 |
| J&J Sensar AR40e | 118.7 | 118.7 ULIB | – | 118.71 | 118.70 | – | 118.40 | 118.7 |
| B&L enVista MX60E | 118.2 | 119.1 mfr | – | 119.15 | – | 119.29 | 119.30 | 119.1 |
| Zeiss CT Asphina 409M | 118.0 | 118.3 ULIB | 118.34 (939) | 118.32 | 118.34 | 118.24 | 117.95 | 118.34 |
| Zeiss CT Lucia 621P | 120.2 | 120.2 mfr | 119.76 (1,178) | – | – | – | – | 119.6 |
| Rayner RayOne EMV | 118.0 US | 118.6 mfr | 118.42 (350) | 118.29 | – | – | 118.60 | 118.1 |
Approximate values from iolcon.org and each calculator’s published constants. US = ultrasound; mfr = manufacturer’s optical value; – = not listed. Check current values before use.
Each calculator uses its own constant for the same lens, so a constant should not be moved from one formula to another. Use the constant made for your formula and biometer, then personalise it with your own outcomes once you have enough eyes. A manufacturer’s nominal value can sit well away from outcomes: for the CT Lucia 621P, 120.2 nominal against 119.76 from 1,178 eyes on IOLcon, about 0.3 D of IOL power.
Where thin-lens formulas fail
Short eyes: each millimetre of lens-position error costs more dioptres, and AL and K alone predict position poorly. Long eyes: hyperopic surprise, which the Wang-Koch adjustment was built to remove. SRK/T’s corneal-height step also misbehaves at very steep or flat K.
The classic rule of thumb
Aristodemou and colleagues compared optimised Hoffer Q, Holladay 1 and SRK/T in 8,108 eyes (2011). It did not include Haigis.
| Axial length | Best of the three | Detail |
|---|---|---|
| 20.0–21.0 mm | Hoffer Q | Lowest MAE; Hoffer Q and Holladay 1 better than SRK/T up to 21.5 mm |
| 22.0–23.5 mm | No difference | All three equivalent |
| 23.5–26.0 mm | Holladay 1 (trend) | Significant in two subgroups |
| ≥ 27 mm | SRK/T | Significantly lowest MAE |
Haigis predicts lens position from measured ACD as well as AL. With all three constants (a0, a1, a2) optimised on a lens-specific outcome dataset it performs well across a wide range, including short steep eyes and long eyes in one 611-eye series; with only a0 optimised it behaves like single-constant formulas, and it still received a Wang-Koch long-eye adjustment. Current formulas such as Kane, Barrett UII and EVO generally match or exceed all of these at both extremes; the classic rules remain useful where only third-generation formulas are available, as on many immersion biometers.
Thick-lens optics
Thick-lens models give the cornea two surfaces and the IOL real thickness, so the lens’s own design starts to matter.
Thick cornea: two surfaces, real thickness
| Eye | Posterior / anterior radius | Basis |
|---|---|---|
| Normal | ≈ 0.82 | Standard ratio built into keratometry |
| After myopic LASIK/PRK | ≈ 0.74 | Measured fall 0.07–0.08 (199 eyes) |
| After hyperopic LASIK | ≈ 0.86–0.90 | Direction documented; value approximate |
| After RK | ≈ 0.89 | Pentacam, 29 eyes |
Without a measured PK, calculations assume the standard ratio. A measured PK or TK replaces that assumption.
Lens shape moves the principal planes
Minus meniscus IOLs act further forward
| −5 D lens shape (front / back surface) | Rear principal plane vs optic centre |
|---|---|
| Biconcave (−2.5 / −2.5 D) | 0.0 mm |
| Meniscus (−6.0 / +1.0 D) | 0.3 mm forward |
| Meniscus (−8.5 / +3.5 D) | 0.5 mm forward |
| Meniscus (−12 / +7 D) | 0.8 mm forward |
Thick-lens optics calculation: acrylic n 1.47 in aqueous n 1.336, centre thickness 0.5 mm. The front principal plane moves forward by a similar amount. If a formula ignores the shape, the lens behaves as if it sat further forward, giving a more hyperopic result than predicted. Literature describes the same principal-plane change and separate constants for negative IOLs.
Low- and high-power designs, and exact ray tracing
Most formulas work in the paraxial approximation, where all rays meet at one focus. Some, such as Olsen’s, use exact ray tracing, which follows real ray paths and shows spherical aberration. It matters for aspheric IOL design, large pupils and corneas reshaped by refractive surgery.
Lens design: more than one constant can carry
A formula has to work for every lens a surgeon might implant: biconvex and asymmetric optics, meniscus designs at low and minus powers, thicker optics at high powers, and plate or loop haptics, with or without a posterior vault. Each changes where the lens sits or where its principal planes lie.
Biconvex
Reference shape
Asymmetric
Principal planes shift
Meniscus
Low and minus powers; planes move outside the lens
Thicker optic
High powers; planes shift with thickness
Plate haptic
Flat; position set by the plate
Loop haptic, posterior vault
Optic sits behind the haptic plane
Loop haptic, no vault
Optic sits in the haptic plane
Side views, light from the left. Schematic, not to scale.
The traditional way to account for all of this is a single lens constant. A constant moves the predicted lens position, so it shifts every power in the same direction. It cannot follow a design that changes at certain powers, such as a separate low-power optic below about 3 D, or a change in optic shape at high powers. Constants are also not standardised: the same lens can carry different constants in different formulas and from different sources.
Attempts to go beyond one constant. Several formulas give a lens more than one number. The Haigis formula can be triple-optimised: its three constants set how the predicted lens position depends on anterior chamber depth and axial length, fitted to outcomes with one lens. The Castrop formula also carries three constants for each lens: one sets how far the lens sits within the space of the crystalline lens, as a fraction of its thickness; one shifts that position; and one shifts the predicted refraction. The Olsen formula takes a different route, combining a single position constant with the lens’s own geometry from the manufacturer and ray tracing through it. Separate constants for minus-power lenses have also been proposed. Each captures more than one constant can, but all still need enough outcomes for each lens, and none follows a design change at a particular power unless the lens geometry is known.
How each calculator handles lens design is often unknown. Some calculators adjust for each lens only through its constant. Others apply lens-specific adjustments when a named lens is selected, as the ZEISS AI IOL Calculation does for its listed lenses. For most calculators, how this is done has not been published. Selecting the exact lens model, rather than entering a constant alone, lets a calculator use whatever lens-specific handling it has. Many calculators probably combine an optimised constant with further lens-specific adjustments that have not been described. Learning such adjustments takes outcomes from many eyes with each lens, most of all at the extremes of power, where few eyes are treated. Cooke and colleagues used 71,183 eyes to describe how corrections for individual lens models change with power. A new lens therefore starts with less certain predictions, which improve as outcomes accumulate. Closer sharing of lens geometry between manufacturers and formula developers, and laboratory studies of lens shape across the power range, could shorten that period.
Ways to study lens behaviour
- Lens geometry from manufacturers. Surface radii, central thickness, refractive index and asphericity for each power, so the lens can be ray traced. Exact, but rarely published.
- The lens in the eye and on the bench. Swept-source OCT after implantation shows where the lens actually sits, so position can be modelled from preoperative measurements and separated from optical error. The same scans image the implanted, hydrated lens itself: its central thickness and, once corrected for refraction, the curvature of its front and back surfaces. Measured across the power range, in the eye or on hydrated lenses on the bench, these show how a design changes shape with power. That adds to what is known of a lens’s geometry where the manufacturer has not published it. Haptic design and vault also change where the optic sits. The average offset is carried by the constant, but how much a haptic is compressed varies with the size of the capsular bag, so the position can differ from eye to eye.
- Postoperative results by lens class. Comparing predicted and actual refraction by lens class, power band and biometer shows what each design does in real eyes, but needs many eyes at the extremes.
No single route is complete. A lens-design effect seen in one should be confirmed by another before it changes a calculation.
Toric IOLs
A toric lens has to cancel the right amount of corneal astigmatism, on the right axis, and stay there.
Neutralising astigmatism
The IOL sits behind the cornea, so it needs more cylinder than the cornea it corrects. The ratio is not fixed: it rises with deeper lens position and steeper corneas, and a fixed ratio such as 1.46 can overstate it.
Toric cylinder and the spherical equivalent
Spectacles
−2.11 D · +1.98 D
SE −0.07 D
Cornea
−2.06 D · +2.02 D
SE −0.02 D
IOL
33.00 D · 27.00 D
SE 30.00 D
- A toric IOL labelled 30.0 D with 6 D of cylinder has meridians of 27.0 D and 33.0 D, which average to 30.0 D at the lens itself.
- At the spectacle plane the two meridians are not mirror images. The IOL sits behind the cornea, so each extra dioptre of IOL power changes the refraction slightly more than the one before.
- The spectacle SE therefore ends up slightly more myopic than with a monofocal lens of the same SE: about 0.02 D at 3 D of IOL cylinder, 0.07 D at 6 D and 0.16 D at 9 D. It grows with the cylinder squared and hardly changes with eye length, lens position or corneal power.
- For most toric powers this is far smaller than a 0.5 D power step. It matters only with very high cylinders.
The shift grows with the square of the cylinder: −0.02 D at 3 D, −0.07 D at 6 D and −0.16 D at 9 D. A longer vertex distance adds slightly: at 6 D, 10, 12 and 14 mm give −0.06, −0.07 and −0.08 D.
The vergence chain, one meridian at a time
| Step | Vergence (n = 1.336, distances in metres) | Flat, 27.00 D | Steep, 33.00 D |
|---|---|---|---|
| 1. Leaving the IOL, focused on the retina | V₁ = n / (AL − ELP) | 84.56 | 84.56 |
| 2. Entering the IOL | V₂ = V₁ − P(IOL) | 57.56 | 51.56 |
| 3. Carried forward to the cornea | V₃ = V₂ / (1 + (ELP/n)·V₂) | 47.02 | 42.94 |
| 4. In air, in front of the cornea | V₄ = V₃ − K | +2.02 | −2.06 |
| 5. Carried to the spectacle plane | R = V₄ / (1 + BVD·V₄) | +1.98 | −2.11 |
| Spectacle SE = (+1.98 − 2.11) / 2 | −0.07 D | ||
Where the toric cylinder sits
- Cylinder placed deeper in the eye corrects less corneal astigmatism per dioptre.
- Posterior toric designs need slightly more cylinder than anterior toric designs; bitoric sits between.
- The difference grows with IOL power and cylinder; the axis is unchanged.
Posterior corneal astigmatism: predicted or measured
- The back surface usually adds against-the-rule astigmatism, about −0.3 D on average, steepest vertically in most eyes.
- Ignoring it over-corrects with-the-rule and under-corrects against-the-rule astigmatism.
- Predicted: Baylor nomogram, Abulafia-Koch regression, or a calculator’s built-in estimate.
- Measured: TK on swept-source OCT, or Scheimpflug posterior keratometry.
Evidence
In normal eyes, predicted posterior astigmatism performs about as well as measured. After LASIK and in keratoconus, where the back surface no longer follows the front, measured values are better.
Surgically induced astigmatism and vector maths
Astigmatism adds as vectors on a double-angle plot. The incision’s own effect (often about 0.1–0.4 D for small incisions) is added to the corneal cylinder to give the cylinder the IOL must correct. Toric calculators differ in how they estimate the back surface and the lens position.
Rotation reduces correction
- About 3.3% of the correction is lost per degree.
- At 10°, about 35% of the cylinder remains; at 30°, all of it; beyond 30°, astigmatism is worse than with no toric.
- The axis of the residual astigmatism also shifts.
Why lenses rotate and what to do
- Most rotation happens in the first hours to days after surgery.
- Long eyes and large capsular bags carry more risk.
- Reposition when the residual cylinder matters, often beyond about 10°. Plan the new axis from the measured refraction with a back-calculator: the best axis is not always the one originally intended, and the expected gain can be small.
Special eyes
Where the standard assumptions break: modified corneas, keratoconus, previous IOLs, no capsular support and silicone oil.
After refractive surgery
Myopic LASIK/PRK
Front flattens, back unchanged: ratio falls to about 0.74. Standard K overestimates power, giving a hyperopic surprise.
Hyperopic LASIK
Centre steepens: ratio rises (about 0.86–0.90, approximate). The error reverses.
Radial keratotomy
Both surfaces flatten, but not in parallel: ratio rises to about 0.89. Irregularity and refractive drift add uncertainty.
Methods use the clinical history, or work without it: Barrett True-K, Haigis-L, EVO with measured PK or TK, and Potvin-Hill, which performed best in eyes with a back-to-front ratio of 0.70 or less in one 2024 study.
Lens choice: myopic LASIK/PRK makes the cornea’s spherical aberration more positive, so an aspheric IOL with more negative spherical aberration offsets more of it. Hyperopic LASIK shifts it the other way, where an aberration-neutral or spherical IOL suits better. Strongly aspheric lenses are more sensitive to decentration and tilt.
Why total keratometry falls below K after myopic LASIK
Standard keratometry measures only the front of the cornea and converts it with a fixed index of 1.3375. That index assumes the back surface keeps a fixed proportion to the front. Total keratometry (TK) adds the measured back surface and the corneal thickness, so it assumes no proportion.
| Untreated | After −6 D | |
|---|---|---|
| Front radius (mm) | 7.70 | 8.70 |
| Front surface power (D, n 1.376) | 48.83 | 43.23 |
| Anterior K (D, keratometric index 1.3375) | 43.83 | 38.81 |
| Back radius (mm) | 6.80 | 6.80 |
| Posterior K, measured (D) | −5.88 | −5.88 |
| Posterior K that K assumes (fixed 0.883 ratio) | −5.88 | −5.21 |
| Central thickness (mm) | 0.50 | 0.43 |
| Total K, thick lens, true indices (D) | 43.05 | 37.43 |
| TK equivalent (D, IOLMaster-style scale) | 43.83 | 38.11 |
| TK − K (D) | 0.00 | −0.70 |
| Back / front radius ratio | 0.883 | 0.782 |
Untreated corneas sit on the TK = K line. Myopic LASIK moves an eye below it: TK − K falls about 0.12 D for each dioptre of correction, from any starting K.
With about ±0.2 D of measurement scatter, corrections up to about −1.7 D can still read TK ≥ K.
- TK below K is strong evidence of previous myopic LASIK/PRK. In a large IOLMaster 700 database, TK − K below zero indicated prior treatment with more than 99.5% likelihood.
- TK at or above K after reported myopic LASIK/PRK has three explanations: a small correction or regression, measurement scatter, or a history that needs checking. The larger the reported correction, the more TK ≥ K points to the history.
- Outcomes: in 3,738 eyes after myopic LASIK/PRK, every modern formula was more accurate with measured posterior cornea or TK than with K alone. In the 11% of eyes with TK − K ≥ 0, formulas using K alone tended to leave eyes more myopic than predicted, by about 0.2 to 0.5 D. With TK the mean error stayed near zero, although accuracy was still slightly lower.
- TK, where available, carries the information K lacks.
Keratoconus
- Steep K misleads lens-position prediction and K overstates power, usually giving a hyperopic surprise.
- Barrett True-K keratoconus with measured posterior cornea or TK, and Kane keratoconus, lead comparisons (TK MAE 0.78 D against 0.83 D with K in one 67-eye series).
- In advanced cones keratometry is least reliable; many surgeons aim slightly myopic.
Pseudophakic eyes: exchange and piggyback
- Exchange: calculate from biometry for the new lens and its expected position, or from the refraction obtained with the old lens and its known power.
- Piggyback: refraction-based methods calculate the power to correct the residual error at the sulcus plane (for example Barrett Rx).
- Optical biometry: choose the pseudophakic setting, and the IOL material where the device asks for it. Biometers that measure each segment separately apply the lens’s own index. The measured ACD is to the existing IOL.
- Ultrasound echoes: the IOL gives a very strong echo followed by a chain of repeat echoes that can be mistaken for the retina. Lowering the gain helps separate them.
- Ultrasound speed: sound travels faster through PMMA and acrylic than through the eye, and slower through silicone. Either set the A-scan for the implanted material, or measure at 1532 m/s and add a correction: a factor for the material multiplied by the central IOL thickness.
- Correct once only: a reading taken with the A-scan already set for the lens is already corrected, so adding the factor again doubles the error.
| IOL material | Add to a 1532 m/s reading | Uncorrected reading |
|---|---|---|
| PMMA | about +0.2 to +0.6 mm | Too short |
| Acrylic | about +0.1 to +0.3 mm | Too short |
| Silicone | about −0.4 to −0.7 mm | Too long |
Approximate ranges for lenses of +10 to +30 D; thicker, higher-power optics need the larger correction. Uncorrected, the error can reach about 2 D. Check the factor and central thickness for the implanted lens.
Aphakia and scleral fixation
- Without a capsule, the optic plane is set by the fixation geometry.
- For intrascleral or sutured fixation, the sclerotomy distance from the limbus and the corneal diameter (WTW) set the optic depth.
- Iris fixation places the optic differently again. For a dislocated IOL, decide on exchange or refixation before calculating.
Silicone oil: measure, then decide whether it stays
- Sound travels at about 980 m/s in 1000 cSt oil against 1532 m/s in vitreous, so ultrasound reads long: correct the vitreous length by 980/1532.
- Optical biometry with the silicone-oil setting is preferred. Pre-oil biometry helps; the fellow eye’s AL reduced accuracy in a 2024 study.
- Oil coming out: calculate for an eye without oil.
- Oil staying: its higher index weakens the IOL’s back surface; add about +3.0 to +3.5 D for a biconvex lens, or use a plano-convex lens with the flat side back.
The formula landscape
From fixed optics to models learned from data, on one map. Internal details of several formulas are proprietary; descriptions follow what has been published.
One map of the model family

Fit the core, extend into the edges

Developing RUI
How RUI is being developed: map where eyes are dense, fit a backbone on the core, add a correction, then score every combination to find the best mix of optics, regression and AI.
From a large dataset to development and validation sets

Where the eyes are

Support in three factors

Three candidate backbones

Which inputs the add-on sees

One network or several

Missing inputs: six ways to train

Scoring every combination
Every backbone, input set, method and missing-input approach is scored the same way on the same held-out eyes. The aim is the mix of optics, regression and AI that is most accurate where eyes are dense and still holds at the edges. That mix is frozen, then scored once on the locked validation set.

From core engine to later versions
The best combination, frozen and scored once on the locked validation set, becomes the core engine. Later engine versions build on it with many smaller optimisations and adjustments. Further regressions add small corrections that improve the fit in particular groups of eyes, without replacing the core.
A formula can only do so much in short eyes. Outcomes vary widely even in clean datasets, because every small error is magnified down the chain: fluctuations in axial length or corneal power readings, and the effects of lens thickness and sex, all move the result more than in an average eye. Toric lens geometry matters more too: with high lens power and high cylinder, the effect of where the toric surface sits is potentially larger for lenses that are not bitoric. High cylinder also makes accurate cylinder measurement, and agreement between measured and predicted total corneal power, more important. Repeat the biometry and check that readings agree, enter every field available, and combine readings from more than one biometer rather than relying on a single measurement.